PIONEER PLUS CATHETER
Report
- Report Number
- 2953200-2010-02206
- Event Type
- Malfunction
- Date Received
- November 12, 2010
- Date of Event
- January 14, 2010
- Report Date
- January 14, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR
- Product Code
- DQY
- PMA / PMN Number
- K081804
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED, AND ITS EVALUATION HAS BEEN COMPLETED. THE NEEDLE WAS EXTENDING OUT AND WAS OUT OF PLANE WITH THE HOUSING. DURING EVALUATION, THE NEEDLE ADVANCEMENT RING WAS ADVANCED BUT THE NEEDLE WAS NOT MOVING AT ALL. WHEN ATTEMPTING TO REMOVE, THE CLIP AT THE LUER CAME OUT WITHOUT ANY RESISTANCE. THE LUER WAS FOUND DETACHED FROM THE HYPO-TUBE. THE ADHESIVE BOND BETWEEN THE LUER AND THE HYPO-TUBE APPEARED BROKEN. THE NEEDLE WAS PULLED OUT THROUGH THE DISTAL END. NO KINKS ON THE NEEDLE OR SHAFT. THERE WAS MINIMAL AMOUNT OF GLUE ON THE GLUE PORT. VERY LITTLE GLUE, CLOSE TO NONE, WAS ON THE HYPO-TUBE. THE MICRO BLAST ON THE HYPO-TUBE SURFACE DOES NOT LOOK ROUGH ENOUGH, WHICH COULD CAUSE THE GLUE NOT TO ADHERE PROPERLY. THE BOND BETWEEN THE LUER AND HYPO-TUBE WAS BROKEN DUE TO INSUFFICIENT APPLICATION OF GLUE.
A PIONEER PLUS CATHETER WAS SELECTED FOR USE IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN UNK LESION. IT WAS REPORTED THAT WHEN PREPPING THE CATHETER, THE NEEDLE WOULD NOT RETRACT. A SECOND PIONEER PLUS CATHETER WAS USED TO COMPLETED THE CASE SUCCESSFULLY. NO CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PIONEER PLUS CATHETER | DQY | MEDTRONIC CARDIOVASCULAR | NA | 662923 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |