- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- September 17, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS RESULTS REVEALED THAT THE HELIX WAS DISENGAGED FROM THE HELICAL CHANNEL. BLOOD WAS PRESENT IN/ON THE HELIX MECHANISM AND ITS SLEEVE HEAD. THE DEFIB CONDUCTOR WAS DISTORTED. THE INNER TUBING WAS KINKED/BUCKLED. THE TIP SEAL WAS OBSERVED TO BE OUT OF POSITION.
IT WAS REPORTED THAT A MODEL 6947 LEAD IMPLANT WAS ATTEMPTED, BUT THE HELIX WOULD NOT EXTEND. THE LEAD WAS NOT USED AND WAS REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Other |