FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 1922661 · Received December 13, 2010

Report

Report Number
2649622-2010-13024
Event Type
Malfunction
Date Received
December 13, 2010
Date of Event
September 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES WERE FOUND. THE FULL LEAD WAS RETURNED AND ANALYZED. ALL CONDUCTORS HAD BLOOD/BODY FLUID, NOT OBSTRUCTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD WAS ATTEMPTED TO BE IMPLANTED, BUT WAS NOT USED AS A STABLE LEAD POSITION COULD NOT BE OBTAINED. A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196 ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other 4076 IMPLANTABLE PACING LEAD