FDA Adverse Event Injury Summary report: N

SIGMA 200 D

MDR report key: 1922648 · Received December 13, 2010

Report

Report Number
6000094-2010-02217
Event Type
Injury
Date Received
December 13, 2010
Date of Event
September 13, 2010
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S2
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) PRELIMINARY ANALYSIS REVEALED NO PACING OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A FOLLOW-UP APPOINTMENT THERE WAS NO TELEMETRY WITH THE DEVICE. THE DEVICE WAS REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA 200 D IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. SD203 ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 Hospitalization| R