FDA Adverse Event Malfunction Summary report: N

COOL PATH DUO 7F, 1304-CP2-7-1.5(5)-L-TE4BE1AB

MDR report key: 1922597 · Received November 19, 2010

Report

Report Number
2030404-2010-00235
Event Type
Malfunction
Date Received
November 19, 2010
Date of Event
October 25, 2010
Report Date
October 25, 2010
Manufacturer
ST. JUDE MEDICAL, IRVINE
Product Code
OAD
PMA / PMN Number
P060019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATIONS HAVE CONFIRMED A LEAK IN THE HANDLE OF THE CATHETER IS CAUSED BY THE LUMEN BREAKING AT THE EDGE OF THE CATHETER HANDLE IN THE PROTECTING TUBE. REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THIS LOT MET MFG REQUIREMENTS PRIOR TO SHIPMENT. A QUALITY IMPROVEMENT PROJECT WAS INITIATED TO ADDRESS THIS ISSUE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THERE WAS A HANDLE LEAK. FORTY FIVE MINUTES INTO THE PROCEDURE (AF ABLATION) THE CATHETER STARTED TO LEAK ON THE PROXIMAL PORTION OF THE CATHETER HANDLE. THE CATHETER WAS REMOVED FROM THE PATIENT WITHOUT ANY ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COOL PATH DUO 7F, 1304-CP2-7-1.5(5)-L-TE4BE1AB NONE OAD ST. JUDE MEDICAL, IRVINE 83563 K28975

Patients

Seq Age Sex Outcome Treatment
1 UNK