FDA Adverse Event Injury Summary report: N

PINNACLE MTL INS NEUT36IDX58OD

MDR report key: 19225022 · Received May 1, 2024

Report

Report Number
1818910-2024-09636
Event Type
Injury
Date Received
May 1, 2024
Date of Event
April 22, 2024
Report Date
May 1, 2024
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
KWA
PMA / PMN Number
K003523
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, "REVISION THA FOR REPEATED HIP DISLOCATION, EXPECTED CERAMIC ON CERAMIC, ON OPENING REVEALED METAL ON METAL". THE PRODUCT WAS NOT RETURNED TO DEPUY SYNTHES, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. SEE ATTACHMENT (IMG_4711, IMG_4712, IMG_4713 AND IMG_4714). PHOTO ANALYSIS REVEALED NOTHING INDICATIVE OF A DEVICE NONCONFORMANCE. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE OBSERVED CONDITION OF THE PINNACLE MTL INS NEUT36IDX58OD WOULD NOT CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 2192263 NUMBER, AND NO NON-CONFORMANCES/MANUFACTURING IRREGULARITIES WERE IDENTIFIED.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

REVISION THA FOR REPEATED HIP DISLOCATION, EXPECTED CERAMIC ON CERAMIC, ON OPENING REVEALED METAL ON METAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2160738 PINNACLE MTL INS NEUT36IDX58OD HIP METAL ACETABULAR LINERS KWA DEPUY ORTHOPAEDICS INC US 2192263

Patients

Seq Age Sex Outcome Treatment
1 90 YR Male Required Intervention ARTICULEZE M HEAD 36MM +8.5