FDA Adverse Event Malfunction Summary report: N

ATTAIN OTW

MDR report key: 1922499 · Received December 13, 2010

Report

Report Number
2649622-2010-12970
Event Type
Malfunction
Date Received
December 13, 2010
Date of Event
September 2, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES WERE FOUND. FURTHER TESTING REVEALED BLOOD ON THE DISTAL CONDUCTOR AT THE TIP TO RING SPACER. THE FULL LEAD WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN WAS UNABLE TO TRACK THE LEAD AROUND A TIGHT BEND IN THE CORONARY SINUS. THE LEAD WAS REMOVED AND A 4196 WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194 ASKU

Patients

Seq Age Sex Outcome Treatment
1 89 YR Other 4574 IMPLANTABLE PACING LEAD| 4074 IMPLANTABLE PACING LEAD| 5360 IMPLANTABLE PULSE GENERATOR