FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 1922445 · Received December 13, 2010

Report

Report Number
2649622-2010-12938
Event Type
Injury
Date Received
December 13, 2010
Date of Event
September 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. FURTHER TESTING REVEALED BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY AND BLOOD IN/ON THE HELIX/LOBE MECHANISM. BLOOD IN THE LOBES WAS OBSERVED. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. FURTHER TESTING REVEALED THAT THE LEAD WAS STRETCHED, AND THE LEAD APPEARED DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AT IMPLANT, THE RIGHT VENTRICULAR LEAD DISSECTED THE SUPERIOR VENA CAVA DURING THE IMPLANT, AND THE VEIN WAS TOO SMALL TO PASS THE LEAD. THE RIGHT VENTRICULAR LEAD WAS NOT USED AND THE CHRONIC RIGHT VENTRICULAR LEAD REMAINS IN USE. IT WAS ALSO REPORTED THAT AT IMPLANT, THREE DIFFERENT LEFT VENTRICULAR LEADS WERE ATTEMPTED, BUT ALL CAUSED DIAPHRAGMATIC STIMULATION. THE THREE LEFT VENTRICULAR LEADS WERE NOT USED, AND A DIFFERENT LEAD WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194 ASKU

Patients

Seq Age Sex Outcome Treatment
1 83 YR Hospitalization| L| O| R 5524M IMPLANTABLE PACING LEAD| 5024M IMPLANTABLE PACING LEAD