FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1922435 · Received December 13, 2010

Report

Report Number
2649622-2010-12930
Event Type
Injury
Date Received
December 13, 2010
Date of Event
September 16, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND; PARTIAL LEAD, IN SEGMENTS, RETURNED AND ANALYZED. BLOOD/BODY FLUID WAS NOTED ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED) AND IN/ON THE HELIX MECHANISM. THE OUTER INSULATION WAS BREACHED CUT, THE HELIX WAS DISTORTED/BENT, AND THE LEAD WAS STRETCHED; LEAD DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HELIX REMAINED EXTENDED WHEN ATTEMPTING TO REPOSITION THE LEAD DURING THE IMPLANT PROCEDURE. ATTEMPTS TO EXTRACT THE LEAD WERE UNSUCCESSFUL AND THE PHYSICIAN REMOVED IT FEMORALLY. THE LEAD WAS NOT USED. WHEN THE LEAD WAS EXTRACTED, TISSUE WAS NOTED ON THE HELIX. A DIFFERENT LEAD WAS PLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention