FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 1922349 · Received December 13, 2010

Report

Report Number
6000144-2010-05932
Event Type
Injury
Date Received
December 13, 2010
Date of Event
September 23, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S031
Removal / Correction Number
Z-0126-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE HAD EARLY ERI (ELECTIVE REPLACEMENT INDICATORS). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD