FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1922346 · Received December 13, 2010

Report

Report Number
2649622-2010-12877
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND; FULL LEAD WAS RETURNED AND ANALYZED. BLOOD IN/ON HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD WAS ABANDONED DUE TO A HIGH/UNSTABLE THRESHOLD, AND APPARENT LEAD CONDUCTOR FRACTURE BASED ON HIGH IMPEDANCE (GREATER THAN 3000 OHMS). AN ATTEMPT WAS MADE TO IMPLANT ANOTHER ATRIAL LEAD, BUT THE LEAD WAS TOO SHORT. ANOTHER NEW ATRIAL LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 82 YR Hospitalization| R 6949 IMPLANTABLE TACHY LEAD| 7278 IMPLANTABLE PACEMAKER/CARDIO/DEFIB