FDA Adverse Event
Injury
Summary report: N
ATTAIN OTW
MDR report key: 1922280
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-12844
- Event Type
- Injury
- Date Received
- December 13, 2010
- Date of Event
- September 10, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P010015/S012
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD WAS INCREASING IN IMPEDANCE AND WAS LOSING CAPTURE IN THE LV TIP TO RV COIL CONFIGURATION. THE PACING POLARITY WAS CHANGED TO RING TO COIL AND HIGHER IMPEDANCE AND HIGH THRESHOLDS WERE NOTED. THE LEAD IS STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4194 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Required Intervention | 4076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |