FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 1922280 · Received December 13, 2010

Report

Report Number
2649622-2010-12844
Event Type
Injury
Date Received
December 13, 2010
Date of Event
September 10, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD WAS INCREASING IN IMPEDANCE AND WAS LOSING CAPTURE IN THE LV TIP TO RV COIL CONFIGURATION. THE PACING POLARITY WAS CHANGED TO RING TO COIL AND HIGHER IMPEDANCE AND HIGH THRESHOLDS WERE NOTED. THE LEAD IS STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194 ASKU

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention 4076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB