FDA Adverse Event Malfunction Summary report: N

REVEAL DX

MDR report key: 1922278 · Received December 13, 2010

Report

Report Number
6000144-2010-05909
Event Type
Malfunction
Date Received
December 13, 2010
Date of Event
September 1, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DSI
PMA / PMN Number
K071655
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS OVERSENSING, BY MEANS OF DOUBLE COUNTING THE QRS COMPLEX. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REVEAL DX IMPLANTABLE LOOP RECORDER DSI MEDTRONIC MED REL, INC. 9528 ASKU

Patients

Seq Age Sex Outcome Treatment
1 71 YR Other