FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 1922249 · Received December 13, 2010

Report

Report Number
2649622-2010-12829
Event Type
Injury
Date Received
December 13, 2010
Date of Event
July 21, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD EXPERIENCED LOSS OF CAPTURE. UPON FURTHER REVIEW, IT WAS FOUND THAT THE ATRIAL, LEFT VENTRICULAR AND RIGHT VENTRICULAR LEADS WERE DISLODGED, POSSIBLY DUE TO TWIDDLER'S SYNDROME. THE LEFT VENTRICULAR LEAD WAS EXPLANTED AND REPLACED. THE ATRIAL AND RIGHT VENTRICULAR LEADS WERE REPOSITIONED, AND REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE S IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6935 ASKU

Patients

Seq Age Sex Outcome Treatment
1 85 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB