FDA Adverse Event
Malfunction
Summary report: N
CAPSURE SENSE
MDR report key: 1922214
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-12803
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- September 9, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P830061/S034
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES WERE FOUND, HOWEVER THE PROXIMAL CONDUCTOR WAS DISTORTED AND HAD BLOOD/BODY FLUID (NOT OBSTRUCTED); THE OUTER INSULATION WAS BREACHED CUT; AND THERE WAS APPARENT EXPLANT DAMAGE. THE FULL LEAD WAS RETURNED AND ANALYZED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE LEAD DISLODGED. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SENSE | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4074 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 88 YR | Other |