FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 1922197 · Received December 13, 2010

Report

Report Number
2649622-2010-12794
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS EXHIBITING A RISING PACING CAPTURE THRESHOLD, MARGINAL SENSING, AND LOWERED IMPEDANCE. IT WAS ALSO NOTED THAT THE UNIPOLAR IMPEDANCE WAS GREATER THAN THE BIPOLAR IMPEDANCE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4068 ASKU

Patients

Seq Age Sex Outcome Treatment
1 51 YR Hospitalization| R 4068 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR