FDA Adverse Event Injury Summary report: N

ATTAIN LV

MDR report key: 1922156 · Received December 13, 2010

Report

Report Number
2649622-2010-12778
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD FRACTURED. HIGH THRESHOLDS WERE OBSERVED ON BOTH THE RIGHT ATRIAL AND LEFT VENTRICULAR LEADS. THE RIGHT ATRIAL LEAD WAS CAPPED AND REPLACED. THE LEFT VENTRICULAR LEAD WAS REMOVED AND REPLACED. IT WAS ALSO REPORTED THAT THE DEVICE HAD LOWER LONGEVITY DUE TO HIGH OUTPUT. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN LV IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 2187 ASKU

Patients

Seq Age Sex Outcome Treatment
1 58 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD