FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1922139 · Received December 13, 2010

Report

Report Number
2649622-2010-12766
Event Type
Malfunction
Date Received
December 13, 2010
Date of Event
September 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES WERE FOUND; THE FULL LEAD WAS RETURNED FOR ANALYSIS. BLOOD WAS OBSERVED IN/ON THE HELIX MECHANISM AND THE SLEEVEHEAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE LEAD HAD UNACCEPTABLE THRESHOLDS OF GREATER THAN 2 VOLTS AND THE PHYSICIAN DID NOT LIKE THE WAY THE CONNECTOR END OF THE LEAD LOOKED. THE LEAD WAS NOT IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 66 YR Other