FDA Adverse Event Injury Summary report: N

SIGMA 300 SR

MDR report key: 1922127 · Received December 13, 2010

Report

Report Number
6000094-2010-02191
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S2
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INTERROGATION OF THE DEVICE WAS UNSUCCESSFUL. NO PACEMAKER FUNCTION WAS SEEN ON THE ELECTROCARDIOGRAM AFTER THE MAGNET WAS PLACED OVER IT. THE PATIENT WAS HOSPITALIZED PRIOR TO THE GENERATOR CHANGE. THE DEVICE WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA 300 SR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. SSR303 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R