FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1922119 · Received December 13, 2010

Report

Report Number
2649622-2010-12756
Event Type
Malfunction
Date Received
December 13, 2010
Date of Event
September 20, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. THE HELIX WAS DISENGAGED FROM HELICAL CHANNEL. THERE WAS BLOOD/BODY FLUID IN THE DISTAL CONDUCTOR; BLOOD IN/ON THE HELIX MECHANISM (SLEEVEHEAD). TIP SEAL OBSERVATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, THE PHYSICIAN DID NOT GET ACCEPTABLE R-WAVE VALUES AT THE FIRST LOCATION. WHEN ANOTHER LOCATION WAS TRIED, THE HELIX WOULD NOT EXTEND AFTER 27 TURNS. THE LEAD WAS REMOVED AND REPLACED WITH ANOTHER LEAD. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 83 YR Other