FUSION OMNI ERCP CATHETER
Report
- Report Number
- 1037905-2010-00632
- Event Type
- Malfunction
- Date Received
- November 18, 2010
- Date of Event
- October 19, 2010
- Report Date
- October 19, 2010
- Manufacturer
- COOK ENDOSCOPY
- Product Code
- GBZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- UNKNOWN
Narratives
THE INFO IN THIS REPORT WAS PROVIDED TO US BY THE COOK FACILITY IN (B)(4) ON BEHALF OF A MEDICAL FACILITY IN (B)(6). EVAL: WE COULD NOT CONFIRM THE REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED TO COOK ENDOSCOPY FOR EVAL. AFTER A REVIEW OF THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED, WE CAN REPORT NO DISCREPANCIES OR ANOMALIES WERE OBSERVED. WE WERE UNABLE TO CONDUCT A SAMPLE TEST FROM THIS LOT BECAUSE NONE OF THE PRODUCT FROM THIS LOT REMAINED IN FINISHED GOODS INVENTORY. A REVIEW OF THE TWELVE MONTH COMPLAINT HISTORY FOR THIS PRODUCT FAMILY WAS CONDUCTED. BASED ON THIS REVIEW, THE LIKELIHOOD OF THIS TYPE OF REPORT IS REMOTE. CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVAL. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. PRIOR TO DISTRIBUTION, ALL FUSION OMNI ERCP CATHETERS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MFG. CORRECTIVE ACTION: NO CORRECTIVE ACTION WARRANTED AT THIS TIME BECAUSE THE REPORT WAS UNABLE TO BE VERIFIED. THE COMPLAINT RISK PRIORITY NUMBER (CRPN) FOR THIS TYPE OF OCCURRENCE HAS BEEN CALCULATED BASED ON THE RATE OF OCCURRENCE AND SEVERITY OF THE OUTCOME. BASED ON THIS ACTIVITY, NO CORRECTIVE ACTION IS WARRANTED AT THIS TIME. THE LIKELIHOOD OF THIS TYPE OF OCCURRENCE IS REMOTE. CUSTOMER QA WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.
DURING AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK ENDOSCOPY FUSION OMNI ERCP CATHETER TO CANNULATE THE COMMON BILE DUCT. THE USER OBSERVED SOMETHING PROTRUDING FROM THE DISTAL END OF THE CATHETER. THE CATHETER WAS REMOVED FROM THE ENDOSCOPE AND ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADD'L PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FUSION OMNI ERCP CATHETER | GBZ, CATHETER, CHOLANGIOGRAPHY | GBZ | COOK ENDOSCOPY | W2881176 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | COOK ENDOSCOPY METII-35-480 WIRE GUIDE| ENDOSCOPE (UNK TYPE) |