FDA Adverse Event
Injury
Summary report: N
SPRINT FIDELIS
MDR report key: 1922083
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-12749
- Event Type
- Injury
- Date Received
- December 13, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S029
- Removal / Correction Number
- Z-0067-0070-2008
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN OPEN CHEST PROCEDURE TO IMPLANT OF AN EPICARDIAL LEAD, IT WAS NOTED THAT THE EXISTING RIGHT VENTRICULAR DEFIBRILLATION LEAD HAD PERFORATED THE HEART. THE LEAD TIP WAS STICKING OUT APPROXIMATELY 1/8TH OF AN INCH, AND HAD LESIONS AROUND IT. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT FIDELIS | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6949 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Required Intervention | 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |