FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART MRX-EMS DEFIBRILLATOR
MDR report key: 1922029
·
Received December 8, 2010
Report
- Report Number
- 1218950-2010-02497
- Event Type
- Malfunction
- Date Received
- December 8, 2010
- Report Date
- November 12, 2010
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CUSTOMER REPORTED THAT THE UNIT WILL NOT START, MAKES BUZZING NOISE. THE UNIT WAS EVALUATED AT PHILIPS. THE SYMPTOM WAS VERIFIED. THE PROCESSOR PCA WAS REPLACED TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE UNIT WILL NOT START, MAKES BUZZING NOISE.