FDA Adverse Event Malfunction Summary report: N

SECURA DR

MDR report key: 1922021 · Received December 13, 2010

Report

Report Number
6000144-2010-05841
Event Type
Malfunction
Date Received
December 13, 2010
Date of Event
September 14, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S114
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THERE WERE NO ANOMALIES FOUND UPON ANALYSIS. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. THERE WAS A SET SCREW IN THE CONNECTOR BORE. EVALUATION SUMMARY (B)(4) NO ANOMALIES FOUND

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) MALFUNCTIONED DURING IMPLANT. THE ICD'S SETSCREW WAS STRIPPED AND THE "WRENCH WILL NOT ENGAGE". ANOTHER DEVICE WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SECURA DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D224DRG ASKU

Patients

Seq Age Sex Outcome Treatment
1 75 YR Other 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD