FDA Adverse Event
Malfunction
Summary report: N
SECURA DR
MDR report key: 1922021
·
Received December 13, 2010
Report
- Report Number
- 6000144-2010-05841
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- September 14, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S114
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THERE WERE NO ANOMALIES FOUND UPON ANALYSIS. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. THERE WAS A SET SCREW IN THE CONNECTOR BORE. EVALUATION SUMMARY (B)(4) NO ANOMALIES FOUND
Description of Event or Problem · 1
IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) MALFUNCTIONED DURING IMPLANT. THE ICD'S SETSCREW WAS STRIPPED AND THE "WRENCH WILL NOT ENGAGE". ANOTHER DEVICE WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SECURA DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D224DRG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Other | 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD |