SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2010-12721
- Event Type
- Injury
- Date Received
- December 13, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE FULL LEAD WAS RETURNED IN SEGMENTS AND NO ANOMALIES WERE FOUND. THE DEFIBRILLATION CONDUCTOR WAS DISTORTED; DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED); BLOOD/BODY FLUID ON OUTER TUBING OVERLAY; INNER TUBING KINKED/BUCKLED; OUTER TUBING OVERLAY MELTED AND COSMETIC ENVIRONMENTAL STRESS CRACKING; OUTER INSULATION TORN; EXPOSED DEFIB COIL HAD WHITE SUBSTANCE; OUTER INSULATION COSMETIC DEPRESSION; BLOOD IN/ON HELIX/LOB MECHANISM. THE LEAD WAS STRETCHED AND HAD APPARENT EXPLANT DAMAGE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THAT THERE WAS HIGH IMPEDANCE AND A POSSIBLE LEAD FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. A PACING THRESHOLD OF 2.5V WAS ALSO REPORTED FOR THE LEAD.
IT WAS REPORTED THAT THERE WAS HIGH IMPEDANCE AND A POSSIBLE LEAD FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |