FDA Adverse Event Malfunction Summary report: N

VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE

MDR report key: 19219001 · Received May 1, 2024

Report

Report Number
9610773-2024-30256
Event Type
Malfunction
Date Received
May 1, 2024
Date of Event
April 3, 2024
Report Date
June 21, 2024
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761075015
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

D9 - 4/5/2024, G4 - K190744, H4 - 6/21/2018. THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND THE CUSTOMER'S ALLEGATION WAS CONFIRMED. THE DEVICE EVALUATION FOUND NO ADDITIONAL FINDINGS. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THAT THE FOLLOWING LED TO THE MALFUNCTION: THE VIDEO CABLE IS BROKEN AND THEREFORE THE IMAGE IS NOT DISPLAYED. A DEVICE HISTORY RECORD REVIEW REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

MULTIPLE FOLLOW UPS WERE MADE TO GATHER ADDTIONAL INFORMATION REGARDING THE EVENT REPORTED , HOWEVER WERE UNSUCCESSFUL AS NO RESPONSE IS RECEIVED FROM THE CUSTOMER. THE EVALUATION OF THE EVENT IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE VIDEO TELESCOPE IMAGE TURNED BLACK DURING SURGERY. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE VIDEO TELESCOPE IMAGE TURNED BLACK DURING SURGERY. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2072924 VIDEO TELESCOPE "ENDOEYE HD II", 5.4 MM, 30°, AUTOCLAVABLE LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES) HET OLYMPUS WINTER & IBE GMBH WA50052A 04042761075015

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown