FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1921873 · Received December 7, 2010

Report

Report Number
2531779-2010-02752
Event Type
Malfunction
Date Received
December 7, 2010
Report Date
November 10, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EZ
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING OBSERVED, THE UP, DOWN, AND OK KEYPAD BUTTONS INTERMITTENTLY RESPONDED TO USER INPUTS DURING TESTING, THE CONTRAST KEYPAD BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE, IT WAS OBSERVED THAT THE UP, DOWN, AND CONTRAST KEY CONTACTS WERE MISALIGNED, IT WAS OBSERVED THAT THE UP, DOWN AND OK KEY CONTACTS BECAME INVERTED WHEN PRESSED AND DID NOT ALWAYS SPRING BACK, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL THE KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE KEYPAD BUTTONS REPORTEDLY UNRESPONSIVE WHEN PRESSED. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1