FDA Adverse Event
Malfunction
Summary report: N
9600
MDR report key: 1921831
·
Received December 8, 2010
Report
- Report Number
- 1720753-2010-04528
- Event Type
- Malfunction
- Date Received
- December 8, 2010
- Date of Event
- November 18, 2010
- Report Date
- December 8, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE VOLTAGES WERE VERIFIED AND THE CIRCUIT BOARDS RESEATED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM INTERMITTENTLY FAILED TO PRODUCE FLUOROSCOPY. NO REPORT OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |