FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 1921831 · Received December 8, 2010

Report

Report Number
1720753-2010-04528
Event Type
Malfunction
Date Received
December 8, 2010
Date of Event
November 18, 2010
Report Date
December 8, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE VOLTAGES WERE VERIFIED AND THE CIRCUIT BOARDS RESEATED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM INTERMITTENTLY FAILED TO PRODUCE FLUOROSCOPY. NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1