FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1921820 · Received December 7, 2010

Report

Report Number
2531779-2010-02758
Event Type
Malfunction
Date Received
December 7, 2010
Report Date
November 8, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT (NO LIFTING OR PEELING OBSERVED; HOWEVER, THE BACKLIGHT BUTTON AND UP ARROW BUTTONS WERE INTERMITTENTLY RESPONDING DURING TESTING, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER THE ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE OK BUTTON WAS SLOW IN ACTIVATING THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND THE PUMP REPORTEDLY HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1