FDA Adverse Event
Injury
Summary report: N
VIRTUOSO DR
MDR report key: 1921765
·
Received December 13, 2010
Report
- Report Number
- 6000144-2010-05829
- Event Type
- Injury
- Date Received
- December 13, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Removal / Correction Number
- Z-0128-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS REPLACED DUE TO EARLY BATTERY DEPLETION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D154AWG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD |