VIRTUOSO DR
Report
- Report Number
- 6000144-2010-05824
- Event Type
- Injury
- Date Received
- December 13, 2010
- Date of Event
- September 8, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Removal / Correction Number
- Z-0128-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE DEVICE LOST TELEMETRY AND FUNCTION DUE TO BATTERY DEPLETION. ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.
IT WAS REPORTED THAT WHEN THE DEVICE WAS NEAR THE ELECTIVE REPLACEMENT INTERVAL, THE PATIENT REFUSED TO HAVE THE DEVICE REPLACED. MONTHS LATER, THE DEVICE COULD NOT BE INTERROGATED DUE TO BATTERY END OF LIFE. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D154AWG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Hospitalization| R | 6945 IMPLANTABLE TACHY LEAD| 5524M IMPLANTABLE PACING LEAD |