FDA Adverse Event Malfunction Summary report: N

ADAPTA DR

MDR report key: 1921728 · Received December 13, 2010

Report

Report Number
6000144-2010-05822
Event Type
Malfunction
Date Received
December 13, 2010
Date of Event
September 10, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NVZ
PMA / PMN Number
P980035/S043
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT THERE WAS SENSING DIFFICULTY AND SIGNIFICANT CROSS TALK/OVERSENSING WERE OBSERVED ON THE VENTRICULAR CHANNEL. A DIFFERENT DEVICE WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADAPTA DR IMPLANTABLE PULSE GENERATOR NVZ MEDTRONIC MED REL, INC. ADDRL1 ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention