CAPSUREFIX NOVUS
MDR report key: 1921648
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-12628
- Event Type
- Injury
- Date Received
- December 13, 2010
- Date of Event
- September 10, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS RESULTS REVEALED THAT THERE WERE NO ANOMALIES FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD PERFORATED THE MYOCARDIUM AND THAT THE PATIENT EXPERIENCED DIAPHRAGMATIC STIMULATION AND CHEST PAIN. THE LEAD WAS ALSO NOTED TO HAVE LOSS OF CAPTURE PRIOR TO REPLACEMENT. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| L| R | 5076 IMPLANTABLE PACING LEAD| ADDR01 IMPLANTABLE PULSE GENERATOR |