FDA Adverse Event Injury Summary report: N

SCREW-IN

MDR report key: 1921581 · Received December 13, 2010

Report

Report Number
2649622-2010-12606
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
K902002
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. IT WAS ALSO NOTED THAT THE DISTAL CONDUCTOR WAS DISTORTED AND THERE WAS APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD HIGH THRESHOLD AND THE PATIENT EXPERIENCED PHRENIC NERVE STIMULATION WHICH REQUIRED THE DEVICE TO BE REPROGRAMMED TO RIGHT VENTRICULAR PACING ONLY. THE LV LEAD WAS CAPPED, PARTIALLY EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SCREW-IN IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5071 ASKU

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD