FDA Adverse Event
Injury
Summary report: N
SCREW-IN
MDR report key: 1921581
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-12606
- Event Type
- Injury
- Date Received
- December 13, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- K902002
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. IT WAS ALSO NOTED THAT THE DISTAL CONDUCTOR WAS DISTORTED AND THERE WAS APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD HIGH THRESHOLD AND THE PATIENT EXPERIENCED PHRENIC NERVE STIMULATION WHICH REQUIRED THE DEVICE TO BE REPROGRAMMED TO RIGHT VENTRICULAR PACING ONLY. THE LV LEAD WAS CAPPED, PARTIALLY EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SCREW-IN | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5071 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD |