FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 1921547 · Received December 13, 2010

Report

Report Number
2649622-2010-12586
Event Type
Injury
Date Received
December 13, 2010
Date of Event
September 8, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. SECONDARY FINDINGS INCLUDED OUTER INSULATION MELTED AND ALL CONDUCTORS BLOOD/BODY FLUID (NOT OBSTRUCTED). FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S LEFT VENTRICULAR (LV) LEAD HAD DISLODGED AND HAD A HIGH THRESHOLD. THE LEAD WAS REMOVED AND RETURNED. A NEW LV LEAD WAS IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194 ASKU

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention 5076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB