FDA Adverse Event Injury Summary report: N

ATTAIN STARFIX

MDR report key: 1921539 · Received December 13, 2010

Report

Report Number
2649622-2010-12580
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P060039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS SUSPECTED OF PREMATURE BATTERY DEPLETION AND WAS EXPLANTED. THE DEVICE WAS UPGRADED TO A CRT SYSTEM. THE FIRST LEFT VENTRICULAR LEAD WAS RETURNED DUE TO BEING UNABLE TO FIT INTO THE PATIENT SMALL AND ACUTE VENOUS PATHWAY. A DIFFERENT LEFT VENTRICULAR LEAD WAS SUCCESSFULLY PLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN STARFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4195 ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD