FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 1921528 · Received December 13, 2010

Report

Report Number
2649622-2010-12575
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) OUTER INSULATION BREACHED EVIRONMENTAL STRESS CRACKING (ESC). THE FULL LEAD WAS RETURNED AND ANALYZED. ALL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), OUTER INSULATION MELTED, OUTER INSULATION COSMETIC ESC.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD INFECTION, INCLUDING POCKET EROSION. THE LEAD WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194 ASKU

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5592 IMPLANTABLE PACING LEAD| 6944 IMPLANTABLE TACHY LEAD