FDA Adverse Event
Injury
Summary report: N
MAXIMO II CRT-D
MDR report key: 1921520
·
Received December 13, 2010
Report
- Report Number
- 6000144-2010-05784
- Event Type
- Injury
- Date Received
- December 13, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S084
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THERE WAS T-WAVE OVERSENSING FOLLOWING BIVENTRICULAR PACING. THERE WAS NO T-WAVE OVERSENSING OF INTRINSIC RHYTHMS. THE DEVICE WAS REPROGRAMMED TO USE A DIFFERENT SENSING VECTOR AND THE DEVICE AND LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAXIMO II CRT-D | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | D284TRK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention | 4076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD |