FDA Adverse Event Injury Summary report: N

MAXIMO II CRT-D

MDR report key: 1921520 · Received December 13, 2010

Report

Report Number
6000144-2010-05784
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S084
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS T-WAVE OVERSENSING FOLLOWING BIVENTRICULAR PACING. THERE WAS NO T-WAVE OVERSENSING OF INTRINSIC RHYTHMS. THE DEVICE WAS REPROGRAMMED TO USE A DIFFERENT SENSING VECTOR AND THE DEVICE AND LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAXIMO II CRT-D IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. D284TRK ASKU

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention 4076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD