FDA Adverse Event Injury Summary report: N

KAPPA 700 DR

MDR report key: 1921484 · Received December 13, 2010

Report

Report Number
6000094-2010-02174
Event Type
Injury
Date Received
December 13, 2010
Date of Event
September 5, 2010
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT CAME TO THE EMERGENCY ROOM (ER) WITH A HEART RATE OF 20 BPM AND NO PACING FROM THEIR DEVICE. THE DEVICE COULD NOT BE INTERROGATED AND THERE WAS NO MAGNET RATE. THE DEVICE WAS EXPLANTED AND RETURNED AND NEW DEVICE WAS IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KAPPA 700 DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. KDR731 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R