FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 11.5MM

MDR report key: 19214774 · Received April 30, 2024

Report

Report Number
0002023141-2024-01389
Event Type
Injury
Date Received
April 30, 2024
Report Date
August 18, 2024
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024020054
PMA / PMN Number
K061410
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). D10: CONCOMITANT MEDICAL PRODUCT AND THERAPY DATES: TSVB11, IMPL TAPERED SCR-V MTX . 3.7MM 3.5MM 11.5MM, LOT NUMBER: 1233148. G4: PREMARKET IDENTIFICATION: K013227.

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1230281. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1230281 , FOR SIMILAR EVENTS AND ONE OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : MEDICAL : BONE LOSS. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR BONE LOSS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

THIS COMPLAINT REFERS TO THE TREATMENT MENTIONED IN CE-44322-2024: DOCTOR REPORTED MASSIVE BONE LOSS WITH INFLAMMATION AT TOOTH SITES 16 AND 15. DOCTOR NOTICED IT FOR A YEAR AND TRIED TO DO THERAPY WITHOUT ANY SUCCESS. PATIENT REFERRED PAIN.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2143567 IMPL TAPERED SCR-V SBM 4. 7MM 4.5MM 11.5MM DENTAL IMPLANT DZE ZIMMER DENTAL 1230281 00889024020054

Patients

Seq Age Sex Outcome Treatment
1 NA Male DENTAL IMPLANT.| DENTAL IMPLANT.