FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX
MDR report key: 1921352
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-12476
- Event Type
- Injury
- Date Received
- December 13, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD HAD AN APPARENT CONDUCTOR FRACTURE. HIGH IMPEDANCE AND BOTH OVERSENSING AND UNDERSENSING WERE ALSO NOTICED ON THE LEAD. THE LEAD WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4068 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Required Intervention | 4024 IMPLANTABLE PACING LEAD| P1501DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB |