ANGLED REAMER SLEEVE, 20°
Report
- Report Number
- 1220246-2024-02615
- Event Type
- Malfunction
- Date Received
- April 30, 2024
- Date of Event
- April 3, 2024
- Report Date
- April 30, 2024
- Manufacturer
- ARTHREX, INC.
- Product Code
- LXH
- UDI-DI
- 00888867343450
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- 501
Narratives
COMPLAINT WAS CONFIRMED. ONE UNPACKAGED AR-9597-20 ANGLED REAMER SLEEVE BATCH NUMBER: 37622043 WAS RECEIVED FOR INVESTIGATION. THE MATING PART AR-9676 BATCH NUMBER: 022339 WAS RECEIVED SEPARATELY FROM THE ANGLED REAMER SLEEVE. UPON VISUAL INVESTIGATION, IT WAS NOTED THAT THE AR-9597-20 HAD WEAR AND TEAR AND SCRATCHES ON THE SLEEVE COLLAR AS WELL AS IN THE INTERNAL DIAMETER AREA ON THE DISTAL SIDE. FUNCTIONAL TESTING WAS PERFORMED WITH THE AR-9676 BATCH NUMBER: 022339 AND IT WAS NOTED THAT THE MATING PART WAS MET WITH FRICTION AND WAS NOT ABLE TO SMOOTHLY ATTACH AND DETACH. THIS FAILURE CAN BE ASSUMED TO BE RELATED TO THE REPORTED FAILURE. A MOST LIKELY CAUSE FOR THE FAILURES FOUND CAN BE ATTRIBUTED TO WEAR AND TEAR DAMAGE INCURRED OVER REPEATED USAGE. REFER TO INVESTIGATION PHOTOS.
ON 04/03/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9597-20 ANGLED REAMER SLEEVE WELDED WITH THE AR-9676 ANGLED REAMER. THIS OCCURRED DURING A REVERSE TOTAL SHOULDER ARTHROPLASTY PROCEDURE ON 04/02/2024 WHILE REAMING THE DEVICES WELDED TOGETHER, CAUSING THE SURGEON'S WRIST TO SUDDENLY TWIST AND MAKING THE REAMING DIFFICULT. THE CASE CARRIED ON AND WAS COMPLETED SUCCESSFULLY WITH A DIFFERENT REAMER SLEEVE AND DRIVESHAFT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 146502 | ANGLED REAMER SLEEVE, 20° | MANUAL INSTR, GENERAL SURGICAL | LXH | ARTHREX, INC. | ANGLED REAMER SLEEVE, 20° | 37622043 | 00888867343450 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |