FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 1921184
·
Received December 9, 2010
Report
- Report Number
- 1720753-2010-04576
- Event Type
- Malfunction
- Date Received
- December 9, 2010
- Date of Event
- November 18, 2010
- Report Date
- December 9, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE CIRCUIT BREAKER WAS RESET, THE TRANSFORMER REPLACED, AND THE MONITOR POWER SUPPLY WAS ADJUSTED DURING THE SVC CALL. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9900 SYS TOUCH SCREEN WOULD NOT WORK AND THE SYS WOULD NOT TURN ON. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |