FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 1921133 · Received December 13, 2010

Report

Report Number
6000144-2010-05701
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S062
Removal / Correction Number
Z-0128-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED RECOMMENDED REPLACEMENT TIME (RRT) EARLY. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154AWG ASKU

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization| R 6940 IMPLANTABLE PACING LEAD| 6945 IMPLANTABLE TACHY LEAD