FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 1921114
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-12370
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- September 9, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT'S RIGHT ATRIAL (RA) PACE/SENSE LEAD HAD CAUSED THE DEVICE TO SWITCH FROM BIPOLAR TO UNIPOLAR PACING DUE TO HIGH PACING IMPEDANCES. IT WAS ALSO REPORTED THAT THERE WAS OVERSENSING. THE PATIENT IS BEING EVALUATED TO SEE IF THE LEAD NEEDS TO BE REPLACED AS THE PATIENT'S NEED FOR PACING IS NO LONGER CLEAR. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |