4 SHOOTER SAEED MULTI-BAND LIGATOR
Report
- Report Number
- 1037905-2010-00591
- Event Type
- Malfunction
- Date Received
- November 11, 2010
- Date of Event
- October 8, 2010
- Report Date
- October 12, 2010
- Manufacturer
- COOK ENDOSCOPY
- Product Code
- MND
- PMA / PMN Number
- K944220
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- UNKNOWN
Narratives
THE INFO IN THIS REPORT WAS PROVIDED TO US BY THE COOK FACILITY IN (B)(6) ON BEHALF OF A MEDICAL FACILITY IN (B)(6). (B)(6). EVAL: WE COULD NOT CONFIRM THE REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED TO COOK ENDOSCOPY FOR EVAL. AFTER A REVIEW OF THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED, WE CAN REPORT NO DISCREPANCIES OR ANOMALIES WERE OBSERVED. WE WERE UNABLE TO CONDUCT A SAMPLE TEST FROM THIS LOT BECAUSE NONE OF THE PRODUCTS FROM THIS LOT REMAINED IN FINISHED GOODS INVENTORY. A REVIEW OF THE 12 MONTH COMPLAINT HISTORY FOR THIS PRODUCT FAMILY WAS CONDUCTED. BASED ON THIS REVIEW, THE LIKELIHOOD OF THIS TYPE OF REPORT IS RARE. CONCLUSIONS: BAND DEPLOYMENT DIFFICULTY CAN OCCUR IF THE ENDOSCOPE ACCESSORY CHANNEL IS COMPROMISED. IN THESE CASES, THE ENDOSCOPE ACCESSORY CHANNEL COLLAPSES, RESTRICTING THE TRIGGER CORD AND PREVENTING PROPER BAND DEPLOYMENT. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING STATEMENT: "THE USE OF AN ENDOSCOPE IN A SOUND STATE OF REPAIR IS A PREREQUISITE FOR A SUCCESSFUL MULTI-BAND LIGATION PROCEDURE." ANOTHER POSSIBLE CONTRIBUTING FACTOR TO BAND DEPLOYMENT DIFFICULTY INCLUDES ALLOWING THE TRIGGER CORD TO BECOME LODGED BETWEEN THE BARREL AND THE DISTAL END OF THE ENDOSCOPE. THIS CAN RESTRICT TRIGGER CORD MOVEMENT AND RESULT IN BAND DEPLOYMENT DIFFICULTY. BAND DEPLOYMENT DIFFICULTY CAN OCCUR IF THE TRIGGER CORD IS NOT PROPERLY SEATED IN THE HANDLE ASSEMBLY. THE INSTRUCTIONS FOR USE ADVISE THE USER THAT THE KNOT MUST BE SEATED INTO THE HOLE OR THE HANDLE WILL NOT FUNCTION PROPERLY. THE INSTRUCTIONS FOR USE DIRECT THE USER TO PLACE THE HANDLE IN THE TWO-WAY POSITION AND LOOSEN THE TRIGGER CORD SLIGHTLY IF THE BAND WILL NOT DEPLOY. THE USER IS THEN INSTRUCTED TO RETURN THE HANDLE TO THE FIRING POSITION AND CONTINUE WITH DEPLOYMENT OF THE BAND(S). THE INSTRUCTIONS FOR USE DIRECT THE USER TO REMOVE THE ENDOSCOPE AND ATTACH A NEW LIGATOR IF MORE BANDS ARE REQUIRED. SLIPPAGE OF THE BAND FROM THE VARICES CAN OCCUR IF THE ENDOSCOPE IS ADVANCED BEYOND THE BANDED SITE. THE INSTRUCTIONS FOR USE ADVISE THE USER THAT PASSING THE ENDOSCOPE OVER A PREVIOUSLY PLACED BAND MAY DISLODGE THE BAND. PRIOR TO DISTRIBUTION, 4 SHOOTER SAEED MULTI-BAND LIGATORS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN IMPLEMENTED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. THE LIKELIHOOD OF THIS TYPE OF OCCURRENCE IS RARE. CUSTOMER QA WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.
DURING AN ESOPHAGEAL BAND LIGATION PROCEDURE, THE PHYSICIAN USED A COOK ENDOSCOPY 4 SHOOTER SAEED MULTI-BAND LIGATOR. FRICTION IN DEPLOYING THE FIRST BAND WAS REPORTEDLY HIGHER THAN NORMAL. THE SECOND BAND COULD ONLY BE DEPLOYED WITH WHAT WAS REPORTED TO BE A HIGHER FORCE THAN NORMAL AND THE BAND SLID OFF. THE BLEEDING WAS STOPPED WITH THE THIRD BAND. THE FOURTH BAND COULD NOT BE RELEASED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADD'L PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 4 SHOOTER SAEED MULTI-BAND LIGATOR | MND, LIGATOR, ESOPHAGEAL | MND | COOK ENDOSCOPY | W2787054 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | OLYMPUS GIF-Q160 GASTROSCOPE |