FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 1921057
·
Received December 9, 2010
Report
- Report Number
- 1720753-2010-04577
- Event Type
- Malfunction
- Date Received
- December 9, 2010
- Date of Event
- November 18, 2010
- Report Date
- December 9, 2010
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR INTERFACE BOARD, FLUORO FUNCTIONS BOARD, AND THE X-RAY TUBE WERE REPLACED DURING THE SVC CALL. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9900 SYSTEM LOCKED UP DURING A CASE. THE 9900 SYSTEM HAD TO BE REBOOTED AND THE PROCEDURE WAS COMPLETED. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |