FDA Adverse Event Malfunction Summary report: N

ECAREMANAGER

MDR report key: 1921028 · Received December 8, 2010

Report

Report Number
1125873-2010-00004
Event Type
Malfunction
Date Received
December 8, 2010
Report Date
December 8, 2010
Manufacturer
PHILIPS VISICU
Product Code
MSX
PMA / PMN Number
K012171
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

(B)(4) MESSAGES SENT BY THE HOSPITAL INFO SYSTEM BUT NOT YET REC'D AND PROCESSED BY ECAREMANAGER AT THE REMOTE MONITORING CENTER WILL BE LOST IF THE REMOTE SERVER PROCESSING THE MESSAGES FAILS OR IS RESTARTED. THE TYPE OF MESSAGES INVOLVED ARE COPIES OF MEDICATION ORDERS, ADMISSION/DISCHARGE/TRANSFER DATA, LABORATORY RESULTS, VITAL SIGN DATA, NURSING FLOWSHEET DATA, AND RESPIRATORY DATA. TYPICALLY, THE MESSAGES DO NOT QUEUE UNLESS THERE IS A PROBLEM WITH THE SERVER. THE USUAL RESOLUTION, HOWEVER, IT TO RESTART THE SERVER. THERE HAVE BEEN NO CUSTOMER COMPLAINTS OR REPORTS OF PT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECAREMANAGER CLINICAL DECISION-SUPPORT SOFTWARE MSX PHILIPS VISICU V3.6

Patients

Seq Age Sex Outcome Treatment
1