FDA Adverse Event
Malfunction
Summary report: N
HANDLES & CONNECTOR WITH DISCHARGE CONTROL FOR INTERNA
MDR report key: 1921016
·
Received December 7, 2010
Report
- Report Number
- 3015876-2010-01338
- Event Type
- Malfunction
- Date Received
- December 7, 2010
- Date of Event
- September 15, 2010
- Report Date
- October 1, 2010
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- LDD
- PMA / PMN Number
- K895379
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): PHYSIO-CONTROL REP EVALUATED THE DEVICE AND VERIFIED THE REPORTED TEAR OF THE RUBBER COVER OVER THE SHOCK BUTTON. A REPLACEMENT INTERNAL PADDLE CABLE ASSEMBLY WAS PROVIDED TO THE CUSTOMER.
Description of Event or Problem · 1
AFTER STEAM STERILIZATION, IT WAS OBSERVED THAT THE RUBBER COVER OVER THE SHOCK BUTTON OF THE INTERNAL HANDLE WAS TORN. THE CUSTOMER HAD USED THE INTERNAL HANDLES 15 TIMES. THERE WAS NO PT USE ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HANDLES & CONNECTOR WITH DISCHARGE CONTROL FOR INTERNA | LDD | PHYSIO-CONTROL, INC | NA | 0904 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |