FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 19210015 · Received April 29, 2024

Report

Report Number
1645337-2024-05109
Event Type
Injury
Date Received
April 29, 2024
Date of Event
April 3, 2024
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000488
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: MATERIAL RUPTURE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON JUNE 17, 2024, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON JUNE 19, 2024, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SMOOTH HPG, 350CC BREAST IMPLANT WAS FOUND TO BE RUPTURED. IN ADDITION, SHELL ABRASION WAS NOTED ON THE EDGES OF THE RUPTURE. THE EVALUATION DETERMINED THAT THE POSSIBLE CAUSE OF THE RUPTURE WAS CONSISTENT WITH NORMAL WEAR. SHELL ABRASION SUGGESTS IN-VIVO FOLDING OR CREASING OF THE DEVICE. THIS MAY BE THE RESULT OF THE FOLLOWING FACTORS: CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE SUCH AS TOO SMALL BREAST POCKET AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. IN SOME CASES, THE BREAST IMPLANTS MAY ALSO WEAR OUT OVER TIME. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 350CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT ELECTED TO UNDERGO BILATERAL BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2024, FOR UNSPECIFIED REASONS. HOWEVER, UPON REMOVAL OF THE IMPLANTS, THE LEFT BREAST IMPLANT WAS FOUND TO BE RUPTURED. SUBSEQUENTLY, THE IMPLANTS WERE REPLACED WITH THE FOLLOWING: (LEFT) 250CC MENTOR MEMORYGEL BREAST IMPLANT CATALOG: 3502504BC LOT: 9980982 SN: (B)(6) AND (RIGHT) 250CC MENTOR MEMORYGEL BREAST IMPLANT CATALOG: 3502504BC LOT: 9982383 SN: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1419947 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 6476054 00081317000488

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female Required Intervention